| Function | |||
| Interactive Interface | LCD, Keypad | ||
| Measurement Range |
Static Pressure: 0~300 mmHg SBP:50~300mmHg, DBP:20~245mmHg; Pulse Rate: 0~240bpm |
||
| Accuracy |
Pressure:3mmHg Pulse Rate:3% or ±3bpm whichever is greater |
||
| Reproducibility | Pressure:≤3mmHg | ||
| Method of Measurement | Oscillometry with linear deflation | ||
| Record Capacity |
ABP-010 & ABP-020 :200 records; ABP-011 & ABP-021 :400 records; ABP-021: 800 records (optional); |
||
| Physical Specification | |||
| Monitor Dimension | 113 x 75 x 26 mm | ||
| Monitor Weight | 168g without battery | ||
| Monitor Ingress Protection | IP22 | ||
| Cuff Size |
M5402BM: arm circumference 20~24cm M5403BM: arm circumference 24~32cm M5404BM: arm circumference 32~38cm M5405BM: arm circumference 38~44cm (optional) |
||
| Collision | Still works in case fall on hard surface from 50mm height in any axial direction | ||
| Operating Environment | |||
| Temperature | 5°C ~ +40°C | ||
| Humidity | 10% ~ 95% RH(no condensed) | ||
| Pressure | 800hPa ~ 1060hPa | ||
| Storage and Transport Environment | |||
| Temperature | -25°C ~ +70°C | ||
| Humidity | ≤95% RH(no condensed) | ||
| Pressure | 560hPa ~ 1060hPa | ||
| Battery | |||
| Type | 2 AA(LR6) alkaline or high capacity Ni-HM batteries | ||
| Supply Voltage | 2.4~3.2V | ||
| Battery Life | >96 hour(measure every 15 minutes) | ||